Microcentrifuge Microcen 23, Orto Alresa

The picture may be different from the article and can include supply, which is not in the standard scope.

Microcentrifuge Microcen 23, Orto Alresa

Order-No
Microcen23
Art #:
Microcen 23
Validity: 31.12.2021
Last Update: 14.06.2021
Microcentrifuge Microcen 23, Orto Alresa

Why a small centrifuge must not be provided of mandatory standards about safety?

The centrifuge Microcen 23 is a compact and practical device which meets the essential requirements that demand a reliable result in a short space of time.

Operator can choose three different rotors for 8x15 ml., 10x15 ml. and 12x5 ml., reaching 8.000 RPM/6.511 xg.

It’s provided of safety systems essentials for a safety processing, with brushless engine maintenance free, unbalance switch off, chamber in stainless steel, lid lock motorized…

Ideal for run a reduced number of samples, it’s undoubtedly the most appropriated choice for emergency labs, due to its reduce the time on sample preparation.

Simplifies sample preparation.

 

  • LED screen shown RPM/RCF (on steps of 100 RPM/10 xg) and time.
  • Intuitive handling through start, stops, short spin and play keys.
  • Rotor list on memory.
  • Timer count up/down, from 0 or at “set RPM/RCF” for test reproducibility.
  • Short spin key with speed adjustable.
  • Acceleration and deceleration control in 2 steps.
  • Automatic open lid.
  • Last values remain on memory.
  • Lid latch lock.
  • Microprocessor controlled.
  • Option: free or locked adjustment of RPM/RCF along the run.
  • Induction motor maintenance free
  • Max. capacity: 10 x 15 ml.
  • Speed up to 6.511 xg/8.000 RPM.
  • Quiet: noise level < 60 dB.
  • Lid locking and holding as well as lid dropping protection.
  • Imbalance switch off.
  • Manual overriding.
  • Chamber of centrifugation in stainless steel.
  • Viewing and calibration port in the lid.
  • Several messages keep the operator informed about the device situation.
  • Guard ring between the chamber of centrifugation and the housing.
  • Forced ventilation which reduces the temperature increase.
EU Directives: 98/79/EC, 2006/95/EC, 2004/108/EC, 2011/65/EU, 2012/19/EU.
Standards: EN 61010-1, EN 61010-2-020, EN 61326-1.
 
RT 246
id
RT 247
id
RT 248
id
  + info + info + info
ROTOR ANGLE FIXED 30° ANGLE FIXED 30° ANGLE FIXED 30°
Max. capacity 8x15 ml. 12x5 ml. 10x15 ml.
RPM 8000 8000 8000
Radius(mm) 91 72 89
RCF Max. (xg) 6511 5152 6368
Braking time (s) 15 15 15

 
more
Technical data
Refrigeration available: No